Spirometry Filters
Our spirometry filters are available for our spirometer SpiroConnect, but also for all other common spirometry and pulmonary function devices.
Since the outbreak of the COVID19 crisis, antibacterial/antiviral spirometry filters have become indispensable when performing spirometry. Bacterial filters are also highly recommended when using disposable sensors, as the disposable sensors do not prevent the spread of potentially infectious aerosols.

Our spirometry filters are designed to be as comfortable as possible, with a modified oval mouthpiece for a better containment of the lips and ridges for better clamping between the teeth. in addition, extra attention was paid to the lowest possible resistance to the air flow.
Neumobit

Neumobit is a rubber mouthpiece that fits on spirometry filters for a more anatomically adapted shape during the test. Neumobit is made from an elastomer and only weighs 10 grams. Thanks to the flaps in the mouthcorners the possibility of air leakage is eliminated..
Neumobit is available separately per 60 pieces or as part of a spirometry filter pack.
Versions
The filters are available in different versions:




Available Models
- 1532: for spirometers from Jaeger, Sensormedics, M.E.C., Geratherm
- 1533: for KOKO spirometers
- 1534: for MedGraphics spirometers
- 1535: for SpiroConnect, MIR, Micromedical, Vitalograph spirometers
- 1536: for spirometers from Cosmed, Medisoft
Spirometry tests
The spirometry filters are suitable for various lung function tests:
- Spirometry, Flow / Volume curves
- Pulmonary Diffusion Capacity
- Pulmonary Volumes by Helium Dilution Method
- Nitrogen Lung Diffusion Test
- Plethysmography
- Respiratory resistance testing
- Compliance
Technical specifications
- BFE* (30L/min.): 99.999980 %
- VFE** (30L/min.): 99.999909 %
- BFE (55L/min.): 99.997649 % (5 minute test)
- VFE (55L/min.): 99.998660 % (5 minute test)
- BFE (750L/min.): 99.502439 % (5 minute test)
- VFE (750L/min.): 99.987339 % (5 minute test)
- Resistance: 94.3 Pa/l/s at 14 l/s (maximum resistance recommended by ATS: 150 Pa/l/s at 14 l/s)
- Dead space: 70 ml
- Suitable for children and adults
* BFE: Bacterial Filtration Efficiency
** VFE: Viral Filtration Efficiency